Optical unit (70 mm・Focused spot type) CL-P070N

CL-P070N - Optical unit (70 mm・Focused spot type)

*Please note that accessories depicted in the image are for illustrative purposes only and may not be included with the product.

  • CE Marking
  • CSA

Specifications

Model

CL-P070N*1

Type

Optical unit

Head

CL-P070 *1

Reference distance

70 mm

Reference measurement range

Measurement range

± 10 mm

Linearity

± 2.2 µm*2

High precision measurement range

Measurement range

± 3.0 mm

Linearity

± 2.0 µm *2

Resolution

0.25 µm

Spot diameter

ø 50 µm

Laser class

Optical unit

Class 1

Sampling cycle

100/200/500/1000 µs (Adjustable 4-stage)

Environmental resistance

Enclosure rating

Head

IP67 (IEC60529)

Ambient light

Target surface illuminance 30,000 lux (Incandescent lamp)

Ambient temperature

0 to 50°C

Relative humidity

85% RH or less (no condensation)

Vibration resistance

Head

10 to 57 Hz, double amplitude 1.5 mm; 2 hours each for X, Y, and Z axes

Optical unit

10 to 57 Hz, double amplitude 0.3 mm; 2 hours each for X, Y, and Z axes

Shock resistance

15G 6 ms

Temperature characteristic

Head

0.005% of F.S. / °C

Optical unit

0.015% of F.S. / °C

Material

Head

Front: SUS Rear: Aluminium

Optical unit

Polycarbonate

Weight

Head

Approx. 280 g

Optical unit

Approx. 1600 g

*1 Sensor head and optical unit are a matched pair. Not cross compatible.
*2 Value measured in displacement mode with KEYENCE reference workpiece (mirrored surface).

Data Sheet (PDF) Other Models