Confocal Displacement Sensor
CL-3000 series
Optical unit (Explosion-proof 15 mm model) CL-S015EXN
*Please note that accessories depicted in the image are for illustrative purposes only and may not be included with the product.
Specifications
Model | CL-S015EXN*1 | |||
Type | Optical unit | |||
Head | CL-S015EX *1 | |||
Reference distance | 15 mm | |||
Reference measurement range | Measurement range | ± 1.0 mm | ||
Linearity | ± 0.35 µm*2 | |||
High precision measurement range | Measurement range | − 0.2 mm/− 1.0 mm | ||
Linearity | ± 0.25 µm *2 | |||
Resolution | 0.25 µm*3 | |||
Spot diameter | ø 10 µm | |||
Laser class | Optical unit | Class 1 | ||
Sampling cycle | 100/200/500/1000 µs (Adjustable 4-stage) | |||
Environmental resistance | Enclosure rating | Head | IP50 | |
Ambient light | Target surface illuminance 30,000 lux (Incandescent lamp) | |||
Operating ambient temperature | Head | − 20 to 70°C | ||
Optical unit | 0 to 50°C | |||
Relative humidity | 85% RH or less (no condensation) | |||
Vibration resistance | Head | 10 to 57 Hz, double amplitude 1.5 mm; 2 hours each for X, Y, and Z axes | ||
Optical unit | 10 to 57 Hz, double amplitude 0.3 mm; 2 hours each for X, Y, and Z axes | |||
Shock resistance | 15G 6 ms | |||
Vacuum | Vacuum environments | |||
Explosion-proof rating | Explosion-proof performance | Head | IECEx: Ex op is IIC T6 Ga | |
Optical unit | IECEx: [Ex op is ⅡC T6 Ga] | |||
Zone/Division | Head | IECEx: Zone 0 | ||
Optical unit | Do not install in a hazardous location (associated equipment) | |||
Ambient temperature | Head | − 20℃ ≤ Ta ≤ 70℃ | ||
Optical unit | 0℃ ≤ Ta ≤ 50℃ | |||
Certification number | IECEx: IECEx CSAE 23.0033X | |||
Temperature characteristic | Head | 0.005% of F.S. / °C *4 | ||
Optical unit | 0.015% of F.S. / °C | |||
Material | Head | SUS303/304 | ||
Optical unit | Polycarbonate | |||
Weight | Head | Approx. 350 g | ||
Optical unit | Approx. 1600 g | |||
*1 Sensor head and optical unit are a matched pair. Not cross compatible. |